19
Artis zee
Floor-mounted system for surgical angiography
Laser crosshairs*
Laser crosshairs for FD 30 x 40, integrated into the flat detector housing with tableside operation for simpler and quicker
patient positioning
Class II laser, wavelength 600 – 700 nm (red), < 1 m W output power
Rotatable collimator for 20 x 20 detector
Compact collimator for cardioangiography with rectangular blade and wedge-shaped finger filter
Automatic synchronous rotation of the detector and collimator unit to compensate for image rotation at different
examination positions of the support stand; rotation also possible via remote control
Rotatable collimator for 30 x 40 detector
Angio collimator with rectangular blade, wedge-shaped finger filters for DSA and cardiological applications and graduated
finger filter
Independent rotation and shift of filter blades
Automatic synchronous rotation of the detector and collimator unit to compensate for image rotation at different
examination positions of the support stand; rotation also possible via remote control
X-ray generator
Microprocessor-controlled high-frequency X-ray generator with automatic dose rate control for fluoroscopy and
acquisition
100 kW at 100 kV (DIN 6822)
Nominal power Depends on X-ray tube and focus. See details in pertaining chapter.
SID tracking (automatic tube current adjustment to focus-detector distance)
CAREMATIC automatic X-ray control system for fully automatic calculation and optimization of exposure data based on
fluoroscopic values
Patient transparency monitoring
Monitoring of tube load with data display
kV and mA post-display on image display
Generator control is fully integrated in the system control
Exposure time range 0.5 to 500 ms
Max. continuous power in fluoro mode 3000 W
* Option