Artis zee
Floor-mounted system for surgical angiography
Stand
The Artis zee floor angio system is specifically designed to meet the escalating demands of high-end imaging for
minimally invasive surgical angiography.
C-arm system
Highly flexible and quick positioning
Single joystick for patient-angle oriented C-arm and detector movements
Integrated, ICP computerized collision monitoring (ICP = Intelligent Collision Protection)
Programmable positioning up to 5 system positions,
additional 50 user-definable user positions
and 3 direct positions
Isocenter-to-floor distance 106 cm (41.73“)
Focus-to-isocenter distance 75 cm (29.53“)
Patient coverage 185 cm (72.84“)
Stand rotation motorized programmable positioning from 0° to 35°
Double oblique projections ± 130° LAO/RAO and + 55°/– 45° CRAN/CAUD
at 0° head-end C-arm position;
± 45° LAO/RAO and + 15°/– 45° CRAN/CAUD
at 35° left-side C-arm position
Angulation speed variable up to 25°/s with LAO/RAO and 18°/s with CRAN/CAUD
Variable focus-detector distance approx. 90 cm – 120 cm (35.4“ - 47.24“), speed up to 9 cm/s (3.54“)
Longitudinal C-arm movement motorized up to 15 cm/s (5.09“/s)
Multispace.F* – Maximum positioning flexibility
Additional stand rotation for free positioning of system and table relative to one another, for the following positions, in
addition to others:
Patient access from the left side
Right-side C-arm positioning 30° relative to the longitudinal axis of the patient and double
oblique projections of 58°/65° LAO/RAO and + 45°/– 45° CRAN/CAUD
OR position orthogonal to the longitudinal axis of the patient and double
(Stand left, table rotated) oblique projections of 50°/45° LAO/RAO and + 43°/– 45° CRAN/CAUD
Stand rotation manual from + 60° to – 220°
Orthogonal system control oriented to the longitudinal axis of the patient
Automap*
Automatic stand positioning depending on the reference image selected
Automatic reference image selection depending on the current stand positioning