22
VARIABLE FORMAT <code>
VALID VALUES 0, or 1
DEFAULT VALUE 0
DATA TYPE Integer
DESCRIPTION Determines if data are to be scrolled on the display. ASCII Output only
0 - data are scrolled
1 - data are displayed in place, form atted for a VT100 terminal.
VARIABLE F P S
VALID VALUES 0 to 2147483648
DEFAULT VALUE 100
DATA TYPE long integer
DESCRIPTION Frames per Scan. Sets the number of averaged frames to send to the host
computer when a scan command is issued. Averaged Frames will be sent
to the host computer until the value of FPS is met. At that time the DSA will
exit the scan mode and wait for another command. If a 0 is entered, the
scan will continue until a STOP command is received.
VARIABLE PAGE
VALID VALUES 0 or 1
DEFAULT VALUE 0
DATA TYPE integer
DESCRIPTION W hen set to 1, the DSA will accumulate 10 frames of data before the data
is sent over the network. If FPS is set to a number less than 10, the data will
be sent over the network when FPS is completed. W hen a STOP command
is issued, the frame acquisition will stop on a 10 frame boundary.
NOTE: This variable is active in UDP mode only.
VARIABLE PERIOD
VALID VALUES 125 to 65535 microseconds
DEFAULT VALUE 500
DATA TYPE integer
DESCRIPTION Sets the interval between channel samples. Period is related to the scan
rate, or frequency by the formula:
Comentarios a estos manuales