4
COMMAND CALIBRATE ZERO
COMMAND SYNTAX CALZ
ARGUMENTS None
DESCRIPTION Commands the DSA to perform a quick zero calibration. This operation
produces A/D count values for each pressure channel that is subtracted
from the raw pressure counts before converting to the chosen pressure
units. The data are stored in the ZERO(0-15) and DELTA(0-15)
Configuration Variable Array. CALZ requires approximately 15 seconds to
complete.
RETURNS <nl>
<nl> - end of line
EXAMPLE To update the current ZERO file and correct for any zero drift of the
transducers:
Enter the command:
CALZ
The DSA will measure the zero counts for each channel and update the
Zero and Delta Arrays.
COMMAND CLEAR
COMMAND SYNTAX CLEAR
ARGUMENTS None
DESCRIPTION Commands the DSA to clear any errors that have occurred. The errors are
sent to the client in an ASCII Packet. ASCII Packets are described in a
subsequent sections.
RETURNS <nl>
<nl> - end of line.
EXAMPLE To clear any errors listed in the ERROR Buffer, the following command
would be issued:
CLEAR <CR>
The ERROR buffer will be cleared