D-Link DSA-3200 Manual de usuario Pagina 28

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 52
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 27
24
VARIABLE XSCANTRIG
VALID VALUES 0 or 1
DEFAULT VALUE 0
DATA TYPE integer
DESCRIPTION Sets the External Trigger as the Frame Trigger.
0 = the internal clock is the frame trigger
1 = the external trigger is the frame trigger
VARIABLE Z C
VALID VALUES 0 or 1
DEFAULT VALUE 1
DATA TYPE integer
DESCRIPTION Sets zero correction on or off.
0 = no correction
1 = zero correction
NOTE 1: ZC must be set to 0 for the following conditions:
1. Model is set to 3207
2. The module has gauge transducers installed.
NOTE 2: ZC must be set to 0 for the following conditions:
1. Model is set to 3207
2. Absolute sensors are installed and the CALB command will
not be used.
NOTE 3: ZC must be set to 1 for the following conditions:
1. Model is set to 3217 or 3218
2. Differential transducers are installed
NOTE 4: ZC must be set to 1 for the following conditions:
1. Model is set to 3207
2. Absolute sensors are installed and the CALB command will be
used.
Vista de pagina 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 51 52

Comentarios a estos manuales

Sin comentarios