24
VARIABLE XSCANTRIG
VALID VALUES 0 or 1
DEFAULT VALUE 0
DATA TYPE integer
DESCRIPTION Sets the External Trigger as the Frame Trigger.
0 = the internal clock is the frame trigger
1 = the external trigger is the frame trigger
VARIABLE Z C
VALID VALUES 0 or 1
DEFAULT VALUE 1
DATA TYPE integer
DESCRIPTION Sets zero correction on or off.
0 = no correction
1 = zero correction
NOTE 1: ZC must be set to 0 for the following conditions:
1. Model is set to 3207
2. The module has gauge transducers installed.
NOTE 2: ZC must be set to 0 for the following conditions:
1. Model is set to 3207
2. Absolute sensors are installed and the CALB command will
not be used.
NOTE 3: ZC must be set to 1 for the following conditions:
1. Model is set to 3217 or 3218
2. Differential transducers are installed
NOTE 4: ZC must be set to 1 for the following conditions:
1. Model is set to 3207
2. Absolute sensors are installed and the CALB command will be
used.