23
VARIABLE QPKTS
VALID VALUES 0 or 1
DEFAULT VALUE 0
DATA TYPE integer
DESCRIPTION This switch will control the action the DSA will take when the data buffer is
full.
0 - frames will be discarded when the data buffer is full. The DSA
will continue to scan.
1 - no frames will be lost. The DSA will stop scanning and log an
error if the data buffer is filled.
VARIABLE S I M <code>
VALID VALUES 0, or 1
DEFAULT VALUE 0
DATA TYPE Integer
DESCRIPTION Determines the source of the data samples.
0 - Data are taken from the sensors
1 - Data are generated internally
VARIABLE TIME <code>
VALID VALUES 0, 1, or 2
DEFAULT VALUE 0
DATA TYPE Integer
DESCRIPTION Determines the format of the Time Stamp.
0 - No Time Stamp
1 - Time stamp data are in microseconds
2 - Time stamp data are in milliseconds
VARIABLE UNITSCAN
VALID VALUES see list below
DEFAULT VALUE PSI
DATA TYPE string
DESCRIPTION Unit scan factor. This sets the scan units for the DSA. Setting this value will
also set CVTUNITS. CVTUNITS may be set to a different value, however
UNITSCAN must be set first. The following are the list of units supported
ATM FTH2O KGM2 MH2O OZFT2
BAR GCM2 KIPIN2 MMHG OZIN2
CMHG INHG KNM2 MPA PA
CMH2O INH2O KPA NCM2 PSF
DECIBAR KGCM2 MBAR NM2 PSI
TORR
If a value other than those listed is entered, PSI will be used. It is best to use
uppercase when entering these units.