25
CALIBRATION VARIABLES
VARIABLE A B S
VALID VALUES 0 or 1
DEFAULT VALUE determined by module type
DATA TYPE integer
DESCRIPTION Sets the mode of operation of the module for models 3217 and 3218.
0 = gauge or differential
1 = absolute.
VARIABLE A B S n W here n = 0 - 15 for channels 1 - 16
VALID VALUES 0 or 1
DEFAULT VALUE determined by module type
DATA TYPE integer
DESCRIPTION Sets the mode of operation of the module for model 3207 only.
0 = gauge or differential
1 = absolute.
VARIABLE M A X n Where n = 0 - 15 for channels 1 - 16
VALID VALUES 0 through 500
DEFAULT VALUE 15
DATA TYPE integer
DESCRIPTION Sets the maximum pressure for channel n. This variable is only used if
MODEL is set to 3207. This must be set to zero or to a positive number if
the module is an absolute module.
VARIABLE M I N n Where n = 0 - 15 for channels 1 - 16
VALID VALUES 0 through 500
DEFAULT VALUE 0
DATA TYPE integer
DESCRIPTION Sets the maximum pressure for channel n. This variable is only used if
MODEL is set to 3207. This must be set to zero or to a positive number if
the module is an absolute module.
VARIABLE NEGPTSn Where n = 0 - 15 for channels 1 - 16
VALID VALUES 0 through 8
DEFAULT VALUE 4
DATA TYPE integer
D E S C R IP T ION Se ts th e num b er of nega tive calibratio n poin ts for c hannel n. This must
be set to zero or to a positive number if the module is an absolute module.
This variable is only used if MODEL is set to 3207