VARIABLE NEGPTSL
VALID VALUES 0 through 8
DEFAULT VALUE 4
DATA TYPE integer
DESCRIPTION Sets the number of negative calibration points for channels 1 - 8. This must
be set to zero or to a positive number if the module is an absolute module.
VARIABLE NEGPTSH
VALID VALUES 0 through 8
DEFAULT VALUE 4
DATA TYPE integer
DESCRIPTION Sets the number of negative calibration points for channels 9 - 16. This must
be set to zero or to a positive number if the module is an absolute module.
VARIABLE PMAXL
VALID VALUES any real number
DEFAULT VALUE determined by module type
DATA TYPE float
DESCRIPTION Sets the maximum pressure for channels 1 - 8.
NOTE The module will output 999999 if the measured pressure exceeds this value.
VARIABLE PMAXH
VALID VALUES any real number
DEFAULT VALUE determined by module type
DATA TYPE float
DESCRIPTION Sets the maximum pressure for channels 9 - 16.
NOTE The module will output 999999 if the measured pressure exceeds this value.
VARIABLE PMINL
VALID VALUES any real number
DEFAULT VALUE determined by module type
DATA TYPE float
DESCRIPTION Sets the minimum pressure for channels 1 - 8. This must be set to zero or
to a positive number if the module is an absolute module.
NOTE The module will output -999999 if the measured pressure exceeds this
value.
VARIABLE PMINH
VALID VALUES any real number
DEFAULT VALUE determined by module type
DATA TYPE float
DESCRIPTION Sets the minimum pressure for channels 9 - 16.This must be set to zero or
to a positive number if the module is an absolute module.
NOTE The module will output -999999 if the measured pressure exceeds this
value.